Menu button
Press the MENU button on the comp box to view the Probe Setup menu on the oscilloscope.
Figure 6: Probe Setup menu
Use the menu options on the oscilloscope to change the probe settings.
Ranges
The measurement system has a variety of ranges available for you to select, whether the probe is being used with or
without a tip. This allows for tradeoffs to be made between noise and dynamic range depending on the needs of the
measurement being made.
CAUTION: To avoid damaging the probe, do not exceed the Peak Voltage rating for a given tip or the probe head.
The Maximum Non-Destruct Voltage limit (Peak Voltage) does not increase when the Probe Ranges are changed.
For example, the TIVPMX50X limit of ±250 Vpk is the same for all range settings.
In 4, 5, and 6 Series oscilloscopes, 9 ranges are selectable when Range Mode is set to Manual. The recommended V/div
settings are displayed in the table below. The ranges shown are for the probe SMA input and 1X tip. Multiply the range and
V/div setting by the tip attenuation to get the values for a sensor tip cable.
Table 2: Ranges and 4/5/6 Series MSO Volts/div settings
4/5/6 Series MSO probe ranges
Recommended V/div setting
20 mV
2 mV/div
40 mV
5 mV/div
80 mV
10 mV/div
160 mV
20 mV/div
320 mV
50 mV/div
640 mV
100 mV/div
1.28 V
200 mV/div
2.56 V
500 mV/div
5 V
1 V/div
When using a tip, the label of each sensor tip cable shows the maximum dynamic range and the attenuation factor. For
example, the TIVPMX10X has a 10X attenuation factor and displays a signal with a ±50 V differential voltage. When more
sensitive ranges are selected, dynamic range is limited. Refer to the Linear differential input voltage range in the
specifications table for more information.
Getting started
TIVP Series IsoVuTM Measurement System User Manual
32