Command Descriptions
*SAV (No Query Form)
Saves the instrument settings in an internal nonvolatile memory location.
Nineteen memory locations are available. Switching the power off and on does
not change the settings stored in the registers.
NOTE.
The memory positions 1 to 10 can be protected from the front panel USER
OPT menu. Use the
command to change protected memory
positions to unprotected.
Group
Common
Syntax
*SAV
Arguments
<DECIMAL DATA>
= a number between 1 and 19.
Examples
*SAV
11
SOURce:PULSe:PERiod
The pulse generator time parameters are activated when the output type is
con
fi
gured to pulse using the
PULSe command.
Group
Output
Syntax
SOURce:PULSe:PERiod <Numeric value>
SOURce:PULSe:PERiod?
Arguments
<NUMERIC VALUE>
= a number between 2E-8 (20 ns) and 2 s, in 10 ns increments.
SOURce:PULSe:WIDTh
The pulse generator time parameters are activated when the output type is
con
fi
gured to pulse using the
PULSe command.
Group
Output
FCA3000, FCA3100, MCA3000 Series Programmer Manual
2-99
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...