DSA70000D, DPO70000D/DX/SX,

MSO70000C/DX, DSA70000C,

DPO70000C, and DPO7000C

High Performance Oscilloscopes

ZZZ

Best Practices Manual

*P071298904*

071-2989-04

Summary of Contents for DPO70000C Series

Page 1: ...xx DSA70000D DPO70000D DX SX MSO70000C DX DSA70000C DPO70000C and DPO7000C High Performance Oscilloscopes ZZZ Best Practices Manual P071298904 071 2989 04...

Page 2: ......

Page 3: ...DSA70000D DPO70000D DX SX MSO70000C DX DSA70000C DPO70000C and DPO7000C High Performance Oscilloscopes ZZZ Best Practices Manual xx www tektronix com 071 2989 04...

Page 4: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 5: ...direct static discharge can damage the input circuitry of the instrument To avoid damage follow these techniques to prevent electrostatic discharge to the instrument 1 Discharge the static voltage fr...

Page 6: ...fy that the connectors are free of dirt metal particles scratches and deformations 3 With the connector pointed down use low pressure compressed air at an angle to the connector to clean the connector...

Page 7: ...e connecting and disconnecting cables and adapters The instrument provides a front panel connection for this purpose Figure 1 Wear an antistatic wrist strap Use the supplied connector saver to protect...

Page 8: ...damage caused by rotation of the connectors when installing a connector Use the supplied torque wrench to tighten 8 inch pounds your cable connector to the connector saver CAUTION Failure to use a ba...

Page 9: ...t 3 Insert the cable connector straight into the instrument connector 4 Tighten the connector screws to help maintain the connection and alignment of the connectors Signal path compensation and instru...

Page 10: ...nts Deskew channels Use the Deskew Attenuation control window to compensate for propagation delays of different length probes and to set the input output ratio of any external attenuation or gain betw...

Page 11: ...low the instrument to complete its Start up routine before shutting down the oscilloscope Do not power off the instrument before the Windows system is fully loaded and running as this action will harm...

Page 12: ...k TekConnect inputs 1 5 V relative to the termination bias 30 mA maximum on 1 2 V full scale settings and 8 V on 1 2 V full scale settings MSO70000DX and DPO70000DX Series 1 2 V Full Scale 5 V absolut...

Page 13: ...use an antistatic wrist strap and work at a static approved workstation when you handle the probe See page 1 Preventing electrostatic discharge TriMode probes Dynamic range The differential signal ra...

Page 14: ...el inputs shorted to ground not open the displayed signal should be zero volts If a noticeable zero volt offset is present under the above conditions a Probe Cal operation should be performed See your...

Page 15: ...desoldering either the attachment wires or the damping resistor For best performance and signal integrity keep the lead length between the DUT and the tip as short as possible and the lead lengths th...

Page 16: ...signed to not exceed the minimum bend radius of 2 5 inches Be sure to follow these additional precautions when handling P6700 Series probe cables Never over bend the probe main or tip cables which can...

Page 17: ...nnector To address this issue replace the RF bullets The bullet contacts in the probe body should be replaced every 200 insertion cycles For medium use situations this should be about once a year Remo...

Page 18: ...xtend the holder tangs 2 Insert a new bullet into the tool so that the holder tangs surround the bullet 3 Release the plunger to secure the holder tangs on the bullet 4 Insert the tool into the probe...

Page 19: ...Best practices for using probes MSO70000C DX DPO70000C D DX SX DSA70000C D DPO7000C Best Practices 15...