Installation
5.
Push
Autoset
. The screen should now
display a square wave, approximately
2.5 V at 1 kHz.
NOTE.
For best performance, it is
recommended that you set the Vertical scale
to 500 mV.
If the signal appears but is misshapen,
perform the procedures for compensating
the probe. (See page 23,
If no signal appears, rerun the procedure.
If this does not remedy the situation, have
the instrument serviced by quali
fi
ed service
personnel.
22
MSO4000 and DPO4000 Series Oscilloscopes User Manual