Commands Listed in Alphabetical Order
Arguments
CH1–CH4
speci
fi
es the input channel number, depending on the model of the
oscilloscope.
Examples
TRIGGER:A:RUNT:SOURCE CH4
sets channel 4 as the source for the A pulse
trigger.
TRIGGER:A:RUNT:SOURCE?
might return
:TRIGGER:A:RUNT:SOURCE CH2
indicating that channel 2 is the source for the A pulse trigger.
TRIGger:A:RUNT:WHEn
Sets or returns the type of pulse width the trigger checks for when it detects a runt.
Group
Trigger
Syntax
TRIGger:A:RUNT:WHEn {LESSthan|than|EQual|UNEQual|OCCURS}
TRIGger:A:RUNT:WHEn?
Related Commands
Arguments
OCCURS
argument speci
fi
es a trigger event if a runt of any detectable width occurs.
LESSthan
argument sets the oscilloscope to trigger if the a runt pulse is detected
with width less than the time set by the
command.
than
argument sets the oscilloscope to trigger if the a runt pulse is detected with
width than the time set by the
command.
EQUal
argument sets the oscilloscope to trigger if a runt pulse is detected with
width equal to the time period speci
fi
ed in
within
a ±5% tolerance.
NOTEQual
argument sets the oscilloscope to trigger if a runt pulse is detected with
width greater than or less than (but not equal to) the time period speci
fi
ed in
within a ±5% tolerance.
Examples
TRIGGER:A:RUNT:WHEN THAN
sets the runt trigger to occur when the
oscilloscope detects a runt in a pulse wider than the speci
fi
ed width.
TRIGGER:A:RUNT:WHEN?
might return
:TRIGGER:A:PULSE:RUNT:WHEN
OCCURS
indicating that a runt trigger will occur if the oscilloscope detects a runt
of any detectable width.
2-476
MSO4000 and DPO4000 Series Programmer Manual
Summary of Contents for DPO4000 Series
Page 2: ......
Page 10: ...Table of Contents vi MSO4000 and DPO4000 Series Programmer Manual ...
Page 23: ...Command Syntax MSO4000 and DPO4000 Series Programmer Manual 2 9 ...
Page 24: ...Command Syntax 2 10 MSO4000 and DPO4000 Series Programmer Manual ...
Page 84: ...Command Groups 2 70 MSO4000 and DPO4000 Series Programmer Manual ...
Page 546: ...Commands Listed in Alphabetical Order 2 532 MSO4000 and DPO4000 Series Programmer Manual ...
Page 564: ...Status and Events 3 18 MSO4000 and DPO4000 Series Programmer Manual ...
Page 565: ...Appendix A Character Set MSO4000 and DPO4000 Series Programmer Manual A 1 ...
Page 566: ...Appendix A Character Set A 2 MSO4000 and DPO4000 Series Programmer Manual ...
Page 572: ...Appendix B Reserved Words B 6 MSO4000 and DPO4000 Series Programmer Manual ...