Display Waveform Data
10.
Select
Probe Setup
to de
fi
ne probe
parameters.
On the resulting side-bezel menu:
Select
Voltage
or
Current
to set the
probe type for probes that do not have the
TekProbe II or TekVPI interface.
Use multipurpose knob
a
to set the
attenuation to match the probe.
11.
Select
Deskew
to set the time skew correction
for the channel. Turn multipurpose knob
a
to
adjust the time skew (deskew) correction for
the probe attached to the selected channel.
This shifts acquisition and display of the
waveform left or right, relative to the trigger
time. Use this to compensate for differences in
cable lengths or probe types.
Quick Tips
Using Probes with the TekProbe II and TekVPI Interfaces.
When you attach a probe with the TekProbe II or the TekVPI
interface, the oscilloscope sets the channel sensitivity, coupling, and termination resistance automatically to match the probe
requirements. Tek Probe II probes require use of the TPA-BNC Adapter.
150
DPO3000 Series Oscilloscopes User Manual