Trigger Setup
Trigger Type
Trigger Conditions
Runt
Trigger on a pulse amplitude that crosses one threshold
but fails to cross a second threshold before recrossing the
fi
rst. You can detect positive or negative (or either) runts, or
only those wider than, less than, greater than, equal to, or
not equal to a speci
fi
ed width. Runt triggers are primarily
used on digital signals.
Logic
Trigger when all channels transition to the speci
fi
ed state.
Use multipurpose knob
a
to select a channel. Push the
appropriate side-bezel button to set that channel’s state to
High (H)
,
Low (L)
, or
Don’t Care (X)
.
Use the
Clock
side-bezel button to enable clocked (state)
triggering. You can have at most a single clock channel.
Push the
Clock Edge
bottom bezel button to change the
polarity of the clock edge. Turn off clocked triggering and
return to unclocked (pattern) triggering by selecting the
clock channel and setting it to high, low, or don’t care.
For unclocked triggering, by default, triggering occurs when
the selected condition goes true. You can also select
triggering when the condition goes false, or time-quali
fi
ed
triggering.
118
DPO3000 Series Oscilloscopes User Manual