The first column is the name of the parameter (such as Record Length, Sample Interval, Trigger Point, and Trigger Time), the
second column is the value of the parameter, the third column is the unit associated with each of the parameters, the fourth
column is the time of the waveform sample relative to the trigger point, and the fifth column is the value of the sampled waveform
that was converted from the oscilloscope's internal digitizing levels.
time = sample interval * (index – trigger point )
Where:
sample interval = the Sample Interval from column two
index = the sample number of the sample being calculated
trigger point = the sample number of the Trigger Point
The information in the first to fourth columns is generated only if you select the Include waveform scale factors option and when
you paste or export data.
Deselect the Include waveform scale factors option when you want to analyze and/or plot the data for several waveforms
captured with the same setup on a given instrument. The values in the first to fourth columns are same for all the waveforms. So
you need to export these columns only once.
Set mask test controls
From the Mask menu, select Mask Controls. You can also click the Controls button in the Mask Testing - Pass/Fail Setup or
Mask Testing - Pass/Fail Results control windows.
Overview. Use this control window to run a test and view the results. You can also reset or start the test and return to the Mask
Setup control window from this control window.
To use.
■
Click On/Off to start and stop the test.
■
Click Reset to reset the test.
■
Click the drop-down arrow in the Test Summary field to select a different source. Status, Total Hits, and Failed Wfms are
read only fields.
■
Click Setup to return to the Mask Setup control window.
Behavior. This control window allows a mask test to run on a full screen and still provide access to the start, stop and reset
functions. In addition, you can see the test summary for each input selected.
What do you want to do next?
Oscilloscope reference
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
807
Summary of Contents for DPO2304SX
Page 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Page 2: ......
Page 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Page 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Page 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...