Behavior. The horizontal delay feature is useful when you want to acquire waveform details that are separated from the trigger
event by a significant interval of time.
For example, you can trigger the instrument on a sync pulse that occurs once every 10 ms, and then look at the high-speed
characteristics that occur 6 ms after the pulse.
TIP. A convenient way to use the horizontal delay is to make quick comparisons between two different points in time. Trigger the
instrument on one area of interest and then set the horizontal delay to acquire the other area of interest. You can then toggle the
horizontal delay on and off to compare details at the two points in time.
What do you want to do next?
Learn more about the Horizontal Position controls.
Set the horizontal scale when delay is on
From the Horiz /Acq menu, select Position/Scale to open the
Horizontal Position/Scale control window
Overview.
To use.
■
Click in the Scale entry box and use the multipurpose knob or keypad to change the setting.
Behavior. Use the Scale control to adjust the time base. This control scales all live displayed waveforms at the same time. When
horizontal delay is on, the scale expands or contracts around the horizontal reference point instead of the trigger point.
The scale units depend on the waveform type. In most cases the horizontal units will be time. However, the horizontal units can
also be frequency.
TIP. Use the dedicated front-panel Horizontal Scale knob to change the horizontal scale.
What do you want to do next?
Learn more about the Horizontal Position controls.
Learn about the Horizontal Setup control window.
Learn about the Acquisition Setup control window.
Acquisition mode
From the Horiz /Acq menu, select Horizontal/Acquisition Setup; then open the Acquisition tab.
The Acquisition Mode controls determine the method of acquisition.
Oscilloscope reference
770
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
Summary of Contents for DPO2304SX
Page 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Page 2: ......
Page 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Page 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Page 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...