A-B Trigger Time Interval. The A-B Trigger Time Interval Application also uses trigger hardware to measure the time between
the A-Event and the B-Event in an A-B Sequence trigger. The A- and B-Events may be any combination of trigger types: edge,
glitch, width, and so on. When the reference value is non-zero, the measured value is displayed or queried as a delta from the
reference.
Set up measurement statistics
From the Measure menu, highlight Statistics; then select Statistics Controls from the submenu, or click the Statistics button in the
Measurement Setup window.
Overview
Use the Measurement Statistics control window to display statistical calculations on measurements. The control window contains
controls that determine which statistics are displayed in the measurement window.
To use
■
Click Off to turn the measurement statistics display off.
■
Click Mean to display the value and the mean (µ) of each measurement.
■
Click All to display the statistics for each measurement value.
(µ) Mean
(m) Minimum
(M) Maximum
(
s
) Standard Deviation
(n) Population
■
Click in the Weight entry box, and then use the multipurpose knobs to set the statistics weight.
■
Click Reset to restart the statistics calculations.
TIP. Select the individual statistics directly from the Measure menu Statistics submenu.
Behavior
Measurement values continue as live updates while the statistics accumulate over time. The statistics appear on the main screen
when a control window is open or on the right side of the screen when no control windows are open and there are less than four
measurements active. The color of the statistics match the color of the waveform from which the measurement was taken.
The statistics weight specifies the constant used to calculate the mean and standard deviation; the minimum and maximum
values are unaffected by weighting. A large value produces statistics that vary less over time.
All statistics (mean, minimum, maximum, standard deviation, and population) are always gathered even though the mean may be
the only statistic shown. When measurements are turned on, the acquisitions take longer because measurements and statistic
calculations are done on each acquisition.
What do you want to do next?
Measurement setups
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
287
Summary of Contents for DPO2304SX
Page 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Page 2: ......
Page 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Page 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Page 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...