Start and stop waveform acquisitions
From the Horiz/Acq menu, select Run/Stop.
Overview
Use the Run/Stop control window to start and stop waveform acquisition.
To use
■
Click Run/Stop to toggle waveform acquisition. (You can also press the front panel Run/Stop button.)
■
Click Single Sequence to complete a single acquisition sequence.
■
Click Force Trigger to immediately start acquiring waveform data without a valid trigger event.
Behavior
These controls operate identically to the front-panel controls.
■
Click Run/Stop once to stop waveform acquisition. The Status readout at the top of the graticule changes from Running to
Stopped and shows the number of acquisitions. Click Run/Stop again to resume waveform acquisition.
■
Click Single Sequence to provide a single sequence of acquisitions each time you click the button. When Single Sequence
is enabled, the front-panel single sequence (Single) button is lighted.
■
In Sample, Peak Detect, or Hi Res acquisition modes, the instrument acquires waveform records with the first valid trigger
event and then stops. In Average or Envelope acquisition modes, the instrument makes the specified number of acquisitions
to complete the averaging or enveloping task.
■
In Waveform Database mode, the instrument acquires and displays a waveform that is the accumulation of several
acquisitions. In addition to time and amplitude, the waveform displays a count of the number of times a specific sample point
has been acquired.
■
When you click Single Sequence (or press the Single button) in Equivalent Time mode, the instrument continues to
recognize trigger events and acquires samples until the waveform record is filled.
■
Click Force Trigger once to acquire a waveform record without a trigger event. Forcing a trigger is useful when the input
signal does not contain a valid trigger. The instrument recognizes and acts upon the forced trigger even when you click the
button before the end of the pretrigger holdoff. The button has no effect when acquisitions are stopped.
What do you want to do next?
Go to the Horizontal/Acquisition control window.
Return to the Acquisition Setup control window.
Return to the Trigger Setup control window.
Horizontal and acquisition setups
192
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
Summary of Contents for DPO2304SX
Page 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Page 2: ......
Page 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Page 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Page 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...