Operating requirements
Cooling
Place the instrument on a cart or bench, observing clearance requirements:
■
Top and bottom: 0 cm (0 in)
■
Left and right side: 5.08 cm (2 in)
■
Rear: 0 cm (0 in)
CAUTION. To ensure proper cooling, keep sides of the instrument clear of obstructions.
Environmental requirements
The environmental requirements for your instrument are listed in the following table. For instrument accuracy, ensure that the
instrument has warmed up for 20 minutes and meets the environmental requirements listed in the following table.
Requirement
Description
Temperature (operating)
0 °C to 50 °C (+32 °F to +122 °F)
Humidity (operating)
5% to 90% relative humidity at up to 30 °C (86 °F)
5% to 45% relative humidity above 30 °C (86 °F) up to +50 °C
(122 °F) noncondensing
Altitude (operating)
Up to 3,000 m (9,843 feet)
Power supply requirements
The power supply requirements for your instrument are listed in the following table.
WARNING. To reduce the risk of fire and shock, ensure that the mains supply voltage fluctuations do not exceed 10% of the
operating voltage range.
Source Voltage and Frequency
Power Consumption
100 VAC to 240 VAC, 50/60 Hz
500 W
Cleaning
Inspect the arbitrary waveform generator as often as operating conditions require. Follow these steps to clean the exterior
surface.
WARNING. To avoid personal injury, power off the instrument and disconnect it from line voltage before performing any of the
following procedures.
Install the instrument
AWG70000B Series Installation and Safety Instructions
7
Summary of Contents for AWG70001B
Page 2: ......
Page 6: ......
Page 12: ...Important safety information vi AWG70000B Series Installation and Safety Instructions ...
Page 16: ...Compliance Information x AWG70000B Series Installation and Safety Instructions ...
Page 32: ...Install the instrument 16 AWG70000B Series Installation and Safety Instructions ...