Adjustment Procedures
AWG7000 Series Service Manual
3-17
2.
Click the
Interleave...
button in the Service UI menu to display the following
setup window:
Figure 3-9: Interleave Calibration setup window
3.
Click
Init for Coarse Adjustment
to set the mode to the Coarse Adjustment
settings.
4.
Identify the minimum rise time point for the Interleave (+) Output using the
Measurement function of the sampling oscilloscope while changing the
Ch2
Clock Phase
settings using either the up/down button or the front-panel rotary
knob. You can also enter the value directly using the keyboard.
5.
Click
Set SP (Phase Offset to SP)
at the minimum rise time point to calculate
and set the center point (Search Point: SP). The value will be displayed in the
SP (Search Point) field.
6.
Change the vertical and horizontal scale of the sampling oscilloscope to
50 mV/div and 200 ps/div, respectively.
7.
Click
Init for Fine Adjustment
to set the mode to the Fine Adjustment
settings.
8.
Select
5.0 GHz
in the Sampling Rate field.
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......