![Tektronix AWG7 1B Series Service Manual Download Page 127](http://html1.mh-extra.com/html/tektronix/awg7-1b-series/awg7-1b-series_service-manual_1078422127.webp)
Troubleshooting
Hardware Error
The following table lists the instrument hardware related errors.
Table 4-8: Hardware error codes
Error code
Error log message
Possible defect
on board
Note
300000000
The system cannot
fi
nd the device speci
fi
ed.
MIO
MIO: U100, U200, U220, Y230
300000001
The device is not ready.
N/A
Software
300000002
The AWG register access timeout period has
expired.
MIO, AWG12G
MIO: U200, U300 AWG12G: U1000
MIO-AWG12G Cable
300000003
The time-out was generated while measuring data.
AWG12G
AWG12G: U1000
300000004
The timer for the trigger has stopped.
AWG12G
AWG12G: U1000
300000005
PLL doesn’t lock.
AWG12G,
CLK12G
AWG12G: U1000 CLK12G: U3200,
U3250, Y3260...
300000006
AwgController DLL doesn’t lock.
AWG12G
AWG12G: U1000
300000007
Output board access error Board=No No: Board
number (Start with 0)
OUT10G
OUT10G: U500
300000008
Not enough storage is available to process.
N/A
Software
300000009
Calibration checksum data has an error.
MIO
MIO: U300, U390
300000010
Calibration version unmatched.
MIO
MIO: U300, U390
300000011
YIG tuning has an error.
CLK12G
CLK12G: U3200, U3250, Y3260...
300000012
The Ipc busy timeout period has expired.
AWG12G
AWG12G: U1000
300000013
MemoryController DLL doesn’t lock.
AWG12G
AWG12G: U1000, U400xx
300000014
MemoryController Dynamic Phase Shift - IPC
error.
AWG12G
AWG12G: U1000, U400xx
300000015
MemoryController Dynamic Phase Shift -
Waveform RAM error.
AWG12G
AWG12G: U1000, U400xx
300000016
MemoryController Dynamic Phase Shift - System
Clock error.
AWG12G
AWG12G: U1000, U400xx
300000017
The Diag busy timeout period has expired.
AWG12G
AWG12G: U1000, U400xx
300000019
AWG12G board Calibration checksum data has
an error.
AWG12G
AWG12G: U1090
Software
300000020
AWG12G board Calibration version unmatched.
AWG12G
AWG12G: U1090
Software
300000021
Read Error on Event Input Register.
AWG12G
AWG12G: U1000
300000022
Read Error on Trigger Input Register.
AWG12G
AWG12G: U1000
300000023
Read Error on Trigger Timer Register.
AWG12G
AWG12G: U1000
300000100
Access to EEPROM failed on MIO board.
MIO
MIO: U300, U390
300000101
Access to EEPROM failed on OUTPUT board.
OUT10G
OUT10G: U500, U550
300000102
The relay count is invalid. Board-1
OUT10G
OUT10G: U500, U550
300000103
The relay count is invalid. Board-2
OUT10G
OUT10G: U500, U550
AWG7000B and AWG7000C Series Service Manual
4–73
Summary of Contents for AWG7 1B Series
Page 2: ......
Page 6: ......
Page 16: ...Environmental Considerations x AWG7000B and AWG7000C Series Service Manual...
Page 18: ...Preface xii AWG7000B and AWG7000C Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 22: ...Operating Information 1 2 AWG7000B and AWG7000C Series Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 33: ...Adjustment Procedures...
Page 34: ......
Page 52: ...Adjustment Procedures 3 18 AWG7000B and AWG7000C Series Service Manual...
Page 53: ...Maintenance...
Page 54: ......
Page 130: ...Troubleshooting 4 76 AWG7000B and AWG7000C Series Service Manual...
Page 131: ...Replaceable Parts...
Page 132: ......
Page 136: ...Replaceable Parts Figure 5 1 Cosmetics 5 4 AWG7000B and AWG7000C Series Service Manual...
Page 166: ...Replaceable Parts Figure 5 18 CLK12G module 5 34 AWG7000B and AWG7000C Series Service Manual...
Page 168: ...Replaceable Parts Figure 5 19 AWG12G module 5 36 AWG7000B and AWG7000C Series Service Manual...
Page 170: ...Replaceable Parts Figure 5 20 SMA bracket 5 38 AWG7000B and AWG7000C Series Service Manual...
Page 172: ...Replaceable Parts Figure 5 21 Relay unit 5 40 AWG7000B and AWG7000C Series Service Manual...
Page 174: ...Replaceable Parts Figure 5 22 Rear connectors 5 42 AWG7000B and AWG7000C Series Service Manual...