Performance veri
fi
cation
Select menu
Setting
Operation
Amplitude Unit
Amplitude setting in new
waveform is half of Vpp
New mixed waveform /Sine/Amplitude
Amplitude
DC direct: 5 mV
DC ampli
fi
ed: 50 mV
New mixed waveform /Sine/Amplitude
Output type
DC direct: 5 mV
DC ampli
fi
ed: 50 mV (step 6)
Setting/Analog Ch1/Output type
4.
Verify that each amplitude measurement is within the range speci
fi
ed in the
test record table.
5.
Record the results in the test record.
6.
Repeat steps 3 through 5 for the CH1–, CH2+, and CH2– DC ampli
fi
ed
outputs.
7.
Record the results in the test record.
DC Offset and Vocm tests
This test should be done once in Basic mode and once in Advanced mode to
verify the DC offset and Vocm accuracy of the generator. This test uses a 50
Ω
terminator. It is necessary to know the accuracy of a 50
Ω
terminator in advance
of this test. This accuracy is used for as a calibration factor.
Basic mode
1.
Connect the 50
Ω
terminator to the DMM as shown in the following
fi
gure
and measure the register value.
Figure 5: 50
Ω
terminator accuracy
2.
Calculate the 50
Ω
calibration factor (CF) from the reading value and record
as follows:
CF (Calibration Factor) = 2
/
(1 + 50
Ω
/ Measurement
Ω
)
Measurement (reading of the
DMM)
CF
Ω
Examples
50.50
Ω
1.0050 (= 2 / (1 + 50 / 50.50))
49.62
Ω
0.9962 (= 2 / (1 + 50 / 49.62))
36
AWG4162 Speci
fi
cations and Performance Veri
fi
cation