Performance veri
fi
cation
4.
Set up the generator using the following steps:
Select menu
Setting
Operation
Function
Sine
Sine (front)
Frequency
1.000000 kHz
Frequency/Period (front)
Amplitude Units
V rms The voltage unit should
be matched to that of DMM.
Home > Units > Vrms
Amplitude
35.36 mVrms
Amplitude/High (front)
Channel 1
Output
On
On (front)
5.
Verify that each amplitude measurement is within the range speci
fi
ed in the
test record table.
6.
Record the results in the test record.
7.
Repeat steps 4 through 6 for the CH1–, CH2+, and CH2– outputs.
8.
Record the results in the test record.
Advanced mode
Use the same equipment setup as in the Basic mode test. This will test both DC
direct and DC ampli
fi
ed mode.
1.
Close Basic mode and launch Advanced mode in the generator.
2.
Check that the generator is still connected to the DMM as shown in the
following
fi
gure and that the 50
Ω
terminator is still connected to the
generator’s side output connector.
Figure 4: Amplitude test, Advanced mode
3.
Set up the generator using the following steps:
Select menu
Setting
Operation
New waveform
Sine, 64k points
Home > ArbBuilder> New Std
Set Function to Sine and Total
Number of Points to 64
Sampling rate
64 MS/s
Setting/Timing
AWG4162 Speci
fi
cations and Performance Veri
fi
cation
35