Operating Basics
Synchronous operation
To synchronize multiple instruments, use the Utility menu. The following image
shows the correct setup. Some documentation may also refer to the synchronous
operation as master-slave operation.
1.
Use a BNC cable to connect the front-panel Trigger Output of one instrument
(master) to the Trigger Input of another instrument (slave). The master unit
sends a trigger signal to the slave unit.
2.
Connect the rear panel Ext Ref Input (master) to a 10 MHz signal from
another source.
3.
Connect the rear panel Ext Ref Input (slave) to the 10 MHz signal source.
This signal source controls the clock of the master and slave units.
4.
Push the front-panel Utility button, and then push the System
bezel button of the master unit to display the System menu.
AFG2021 Quick Start User Manual
63