![Tektronix AFG1000 Series Manual Download Page 45](http://html1.mh-extra.com/html/tektronix/afg1000-series/afg1000-series_manual_1077720045.webp)
36
AFG1000 Series Arbitrary Function Generator Programmer Manual
Command Descriptions
When the concurrent copy function is enabled, the FreqLock function is also
enabled automaticlly. Use general knob to adjust frequency (or period) of the two
channels sychronously.
The
[SOURce[1|2]]:FREQuency:CONCurrent?
command returns “0” (off) or “1”
(on).
Group
Source
Syntax
[SOURce[1|2]]:FREQuency:CONCurrent
{ON|OFF|<NR1>}
[SOURce[1|2]]:FREQuency:CONCurrent?
Arguments
ON or <NR1>
≠
0 enables the concurrent copy function.
OFF or <NR1>=0 disables the concurrent copy function.
Returns
<NR1>
Examples
SOURce1:FREQuency:CONCurrent ON
copies the frequency value of CH 1 to CH 2.
[SOURce[1|2]]:FREQuency[:CW|:FIXed]
This command sets or queries the frequency of output waveform for the specified
channel. This command is available when the Run Mode is set to other than
Sweep.
The setting range of output frequency depends on the type of output waveform.
If you change the type of output waveform, it might change the output frequency
because changing waveform types impacts on the setting range of output
frequency. The resolution is 1 μHz or 12 digits. For more information, refer to
the
AFG1000 Series Specifications and Performance Verification Technical
Reference
.
Group
Source
Syntax
[SOURce[1|2]]:FREQuency[:CW|:FIXed]
{<frequency>|MINimum|MAXimum}
[SOURce[1|2]]:FREQuency[:CW|:FIXed]
? {MINimum|MAXimum}
Arguments
<frequency>::=<NRf>[<units>]