Electrical Sampling Modules
7.
Check that the measured fall time (re
fl
ected rise time) for the negative edge
polarity is as follows:
80E04:
≤
35 ps
80E08B:
≤
22 ps
80E10B:
≤
16 ps
8.
Repeat the Setup and Test sections for channel 2.
9.
Disconnect the test setup.
End of Procedure
TDR-System Step Response Aberrations (80E04)
These procedures check those characteristics that relate to the TDR system and are
listed as checked in the
DSA8300 Digital Serial Analyzer Speci
fi
cations
manual.
Equipment required
Terminator, 50
Ω
coaxial terminator, K male (item 28)
Prerequisites
Setup
Figure 26: TDR system step response aberrations setup
1.
Install the module into the left small module compartment of the main
instrument.
2.
Connect a 50
Ω
coaxial terminator to channel 1 as shown in the following
fi
gure.
206
DSA8300 Performance Veri
fi
cation