![Tektronix 80A00 Technical Reference Download Page 217](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946217.webp)
Electrical Sampling Modules
Figure 23: Adapter-loss setup
2.
Perform the following substeps:
a.
Set the signal frequency and the power meter frequency to each of the
frequencies listed in the
Frequency
column of the table. (See Table 9.)
NOTE.
Only perform checks up to the maximum bandwidth of the module
under test.
b.
Turn Trace Hold on the SNA to
On
.
DSA8300 Performance Veri
fi
cation
199