Electrical Sampling Modules
Equipment required
Adapter, 2.4 mm male-to-male (item 1)
Adapter, 2.4 mm female-to-female (item 2)
Attenuator, 10 dB pad, SMA, Femal to Male (item 6)
Cable, special, power/phase stability (item 14)
Power meter (item 24)
Power sensor (item 25)
Synthesized signal generator (item 31)
Prerequisites
Power-Reference
Characterization Setup
Install the test setup to measure the power reference:
NOTE.
Use a torque wrench when connecting the terminators, adapters,
attenuator, and power sensor in this setup. Use 8 ±0.3 inch pounds of torque.
1.
Connect the 10 dB attenuator to the Synthesized signal generator. Connect
the remaining end to the power sensor. Connect the power sensor to the
power meter.
Figure 17: Power-reference setup
Reference
Characterization
Measurements
Follow this procedure to characterize the reference signals:
1.
Set the Synthesized signal generator to 0 dBm.
2.
Set the signal frequency and the power meter frequency to each of the
frequencies listed in the
Frequency
column of the following table.
DSA8300 Performance Veri
fi
cation
185