DSA8300 Performance Veri
fi
cation
Setup
Figure 6: Time interval accuracy test setup
1.
Install the 80E0X module(s) to test starting at the left-most small module
slot as shown in the
fi
gure. Power on the instrument and wait 20 minutes
for warm-up.
2.
Connect the synthesized signal generator to a 1.0 meter cable.
3.
Connect the remaining end of the 1.0 meter cable to a 50
Ω
SMA power
divider. Connect two 0.5 meter cables to the two remaining SMA power
divider connectors.
4.
Connect one of the 0.5 meter cables to Channel
1
of the 80E0X module.
Connect the remaining 0.5 meter cable to the TRIGGER DIRECT INPUT.
5.
Set the synthesized signal generator frequency to 1 GHz.
6.
Set the Synthesized output to
800 mV
p-p
(400 mV displayed)
7.
Set up the instrument as follows:
a.
Push the
Default Setup
front-panel button and click
Yes
.
b.
Select
C1
from the Waveform Selector menu on the controls bar at the
bottom of the display (left corner).
c.
In the
Setup > Mode/Trigger
Setup dialog box:
Select
Direct
for Trigger Source.
d.
Click the
Vert
tab and set the following:
Set the Scale to
100 mV/div
.
Set the Offset to
0
.
152
DSA8300 Performance Veri
fi
cation