Removal and installation
Power PC board
You will need a T-15 Torx driver and a 3/16 inch nut driver.
1.
To access the Power PC board, remove these assemblies:
Trim (all) (See page 29.)
Bottom cover (See page 30.)
Left and Right covers (See page 30.)
CD-RW/DVD and Hard Drive mounting frame (See page 43.)
ATX Assembly (See page 45.)
Power Side board (See Figure 22.)
2.
To remove the Power PC board:
a.
Set the instrument with the bottom down on the work surface and the
top panel facing you.
b.
Disconnect the ribbon cables from the Power PC board.
c.
Remove the ten T-15 Torx drive screws securing the Power PC board to
the chassis support.
d.
Remove the two 3/16 inch securing nuts that attach the GPIB connector
to the rear of the support bracket.
e.
Remove the two 3/16 inch securing nuts that attach the Video out
connector to the rear of the support bracket.
f.
Lift the Power PC board out of the instrument.
3.
To reinstall the Power PC board:
Reverse the steps. Refer to the cable
connections table for information about where to connect cables. (See
page 72,
DSA8300 Digital Serial Analyzer and Modules Service Manual
47