Specification and
Performance Check-5110
PERFORMANCE CHECK
Introduction
This
procedure checks the oscilloscope for measure
ment
accuracy against the tolerances listed as Perfor
mance
Requirements that
appear under Electrical
Characteristics at
the beginning
of this section. If the
instrument
fails to meet the requirements given in this
Performance Check, the Adjustment
procedure (Section 3
in
this manual) should be performed. The Performance
Check can be
used
by an incoming inspection facility to
determine acceptability of
performance. It is
not
necessary
to
remove the instrument
cabinet to perform
this procedure, since all checks are
made from the front
panel.
The
Electrical Characteristics in this section are valid
only
if the oscilloscope has been calibrated at an ambient
temperature b20°
C to +30°
C and is operating at
an ambient
temperature between 0°C to
+50°
C.
PERFORMANCE CHECK
INTERVAL
To
ensure instrument accuracy, check the perfor
mance
of
the oscilloscope every 1000 hours of operation,
or every
6 months
if used infrequently.
TEST
EQUIPMENT
REQUIRED
The following
test
equipment, orequivalent, is required
to perform
a performance check of the oscilloscope. The
test
equipment
performance requirements
listed are the
minimum
required
to verify the performance of the
equipment
under
test. Substitute equipment must meet or
exceed
the
stated requirements. All test equipment is
assumed
to be operating within tolerance.
Tolerances
that
are specified in this Performance
Check
procedure apply to the instrument under test and
do
not
include
test equipment error. Limits
and tolerances
in
this procedure are instrument performance re
quirements only if listed in a Performance Requirements
column
that appears under Electrical Characteristics
at
the beginning of this
section; information given in the
Supplemental
Information column is provided for user
information
only,
and should not be interpreted as perfor
mance requirements.
SPECIAL
TEST
EQUIPMENT
Special
test
equipment is used where necessary to
facilitate
the procedure. Most of
this equipment is
available
from Tektronix,
Inc. and
can be
ordered
through
your
local
Tektronix
Field
Office or representative.
LIST
OF
TEST EQUIPMENT REQUIREMENTS FOR
PERFORMANCE CHECK
Table 2-9
Description
Performance
Requirements
Application
Example
Amplifier Plug-in
Bandwidth, de to 2 MHz;
display
modes,
channel
1
and
dual-trace; deflection
factor,
5 mV
to
5 V/div.
One
required
for all
tests. Two required
for steps
6,
7, 12.
a. TEKTRONIX
5A18N
Dual-Trace Amplifier.
Time-base Plug-in
unit
Sweep rate, at
least
2
µs/div.
Steps
1
through
11,
13, 14.
a. TEKTRONIX
5B10N
Time-Base.
Calibration generator
Amplitude
calibration,
5 mV to
5V; accuracy,
±0.25% into 1
mΩ;
output,
square
wave at approximately
1
kHz.
Steps
8, 10.
a.
TEKTRONIX
PG 506
Calibration Generator
2.
Sine-wave
generator
Frequency, 50 kHz
to 2
MHz; output amplitude,
variable
from
250 mV to
6 V
into 50 Ω.
Steps 9, 11,
12,
13.
a. TEKTRONIX FG 503
Function
Generator2.
'Two
dual-trace
amplifiers
are required
to check vertical alternate
and chop operation. Two Identical amplifiers are required to
check
x-y
phase difference.
"Requires
a TM
500-series power module.
REV A JUL 1979
2-5
Summary of Contents for 5110
Page 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Page 22: ......
Page 34: ......
Page 42: ......
Page 83: ...5110 I43XII3 2134 15 ...
Page 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Page 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Page 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Page 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Page 95: ...DEFLECTION AMPLIFIERs ...
Page 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Page 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Page 114: ......
Page 115: ...5110 OSCILLOSCOPE ...
Page 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Page 117: ...1 ...
Page 121: ...o ...
Page 122: ...S 1 5110 OSCILLOSCOPE ...
Page 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Page 124: ......