Displaying Signals
Figure 3-11: Sample Mode Display
Clocking Acquisitions
Normally, storage acquisitions are clocked internally. You can, however,
supply an external clocking signal through the auxiliary connector on the
side of the instrument. Refer to
Specification
, Appendix B.
Selecting the Acquisition Record Length (1K/4K)
A 1K acquisition consists of 1,024 data points spread across one display
screen. A 4K acquisition consists of 4,096 data points spread across four
screens. To view these additional screens in the 4K mode you must use the
cursor knob.
P I
Step
1: Press the
ACQUISITION
1K/4K button. In the 4K mode an
acquisition window indicator will appear.
The acquisition window indicator (Figure 3-12) displays the relative position
of the cursors, what cursor is active, what part of the acquisition is displayed
(display window indicator), and the trigger point.
n Step 2:
Position the active cursor one way or the other to view the rest
of the acquisition record. (Note that the active cursor and the display
window indicator also move along the acquisition window indicator
Display Window Indicator
Inactive Cursor
Figure 3-12: Acquisition Window Indicator
I
2221A User Manual
3-15
Summary of Contents for 2221A
Page 1: ...User Manual Tektronix 2221A Digital Storage Oscilloscope 070 8156 02 ...
Page 6: ...It Welcome ...
Page 14: ...X 2221A User Manual ...
Page 16: ...xii Product Description ...
Page 26: ...Front Panel CRT Display Figure 2 2 Front Panel View Left Side 2 2 At a Glance ...
Page 48: ...2 24 At a Glance ...
Page 86: ...lopiayiiiij 3 30 In Detail ...
Page 100: ...Saving Waveforms and Setups 3 44 En Detail ...
Page 108: ...Appendix A Options and Accessories Appendices ...
Page 156: ...Appendix C Performance Verification j f i f P P j r fe t E F E J E J E fer 4 52 Appendices ...