background image

 

 

 

www.tek-trol.com   

 

 

 

 

 

 

 

 

                      

20

 

4

 

Electrical Installations 

 

Please cut off power supply before connecting the device 

 

Check the cable model before connecting the cable. 

 

 

Follow the procedure for cable into lead collar: -  

 

 At first, loosen the gland nut on lead collar and take off blind, secondly, put gland nut and 
rubber  ring  on  cable,  make  the  cable  through  lead  collar;  finally,  straighten  out  cable, 
screw gland nut tight to make rubber ring press cable. 

 

When wire stripping, do not damage insulating layer which should be reserved. 

 

Fig 16: TEK-Clamp 1200A-100H Electrical Connections Cables. 

 

Fig 17: Tek-Clamp 1200A-100H Converted Terminal And AC/DC Adaptor 

 

 

Fig 18: TEK-Clamp 1200A-100H Electrical Connections Cables Interface 

 
 

 

Summary of Contents for TEK-C LAMP 1200A-100H

Page 1: ...Technology Solutions T LAMP1200A 100H EK C Instruction Manual Handheld Ultrasonic Flow Meter Document Number IM 1200A 100H www tek trol com...

Page 2: ...tomer Support 796 Tek Drive Crystal Lake IL 60014 USA Tel 1 847 857 6076 COPYRIGHT Tek Trol LLC 2021 No part of this publication may be copied or distributed transmitted transcribed stored in a retrie...

Page 3: ...r Sensors 9 2 4 Dimensional Drawings 10 2 4 1 System Drawings 10 2 4 2 System Description 12 3 Installation 13 3 1 Correct Mounting Point 13 3 1 1 Clamp on Transducer 13 3 1 2 Clamp on mounting bracke...

Page 4: ...ing On 32 6 6 Further Errors and Countermeasures 32 6 7 Further Errors and Countermeasures 33 7 Appendix 34 7 1 Tables Pipe Size 34 7 1 1 Copper Standard Pipe Size Charts 34 7 1 2 PVC Standard Pipe Si...

Page 5: ...humidity condensation or moisture With wet hands never use the instrument Before taking a measurement the device should be stabilised to the ambient temperature important when carrying the device from...

Page 6: ...ation Concerning the Documentation To prevent any injury to the operator or damage to the device it is essential to read the information in this document and the applicable national standard safety in...

Page 7: ...ction cable 2 x Releasable cable tie 1 x Mains adaptor 1 x Contact gel 1 x Tek Trol measuring tape 1 x Carrying case 1 x User manual 1 2 5 3 Optional Accessories S1 Sensor type S1 loose M1 Sensor type...

Page 8: ...environments The following considerations must be observed in selecting a location for the meter The ambient operating temperature range is 22 F 30 C to 176 F 80 C Do not expose the meter to corrosiv...

Page 9: ...ra pure liquids chemicals 2 2 Measuring Principle When the ultrasonic wave is transmitted through the flowing liquid the difference between the upstream and downstream transit time travel time or time...

Page 10: ...2 Resolution 0 0003 ft s Measurement Period 0 999 seconds user configurable Measurement Principle Transit Time Ultrasonic Flow Meter Media All liquids with solids of 5 and a flow of 0 19pm Flow Units...

Page 11: ...il lubricating oil diesel oil fuel oil etc Chemicals alcohol etc Plant Effluent Beverage Ultra pure liquids Process Temperature 22 F to 320 F 30 C to 160 C Pipe Material Steel Stainless Steel Cast Iro...

Page 12: ...36 x1 77 x1 77 7 87 x0 98 x0 98 11 02 x1 57 x1 57 Weight 75 g 260 g 250 g 1080 g 2 4 Dimensional Drawings 2 4 1 System Drawings Fig 2 Top View of 1200A 100H Handheld Ultrasonic Flow Meter Fig 3 Front...

Page 13: ...ol com 11 Fig 4 Bottom View of 1200A 100H Handheld Ultrasonic Flow Meter Fig 5 Cables of 1200A 100H Handheld Ultrasonic Flow Meter Fig 6 Mains Adaptor of 1200A 100H Handheld Ultrasonic Flow Meter Main...

Page 14: ...www tek trol com 12 2 4 2 System Description Fig 7 System Description...

Page 15: ...tallation at wrong point may influence measuring accuracy or flow meter service life or even damage the flow meter The axis for measuring electrode shall be approximate to horizontal direction Tek Cla...

Page 16: ...e of flow meter Magnetic flow meter system should be connected to the ground There must be a straight pipe run no less than 5DN at flow meter upstream and 2DN at flow meter downstream DN is the intern...

Page 17: ...there must be no entrapped air air bubbles in the liquid The pipes can either run vertically or horizontally A straight flow calming section before and behind the measuring point should be considered...

Page 18: ...c waves The ultrasonic waves pass through the pipe walls and the liquid allows conclusions about the flow velocity As the transit time of the ultrasonic pulses is very short the sensors should be inst...

Page 19: ...ata must be entered The outer diameter of the pipe M11 Material thickness of the pipe M12 Material of the pipe M14 The liner of the pipe M16 Type of liquid M20 Type of sensors connected M23 The mounti...

Page 20: ...ol com 18 Fig 11 V Method Installation Top View 3 3 2 2 Z Method Installation Z method is commonly used when the pipe diameter is above 8 200mm Fig 12 Z Method Installation Fig 13 Z Method Installatio...

Page 21: ...www tek trol com 19 3 3 2 3 W method Installation W method is usually used on plastic pipes with a diameter from 2 15mm to 50mm Fig 14 W Method Installation Fig 15 W Method Installation Top View...

Page 22: ...d take off blind secondly put gland nut and rubber ring on cable make the cable through lead collar finally straighten out cable screw gland nut tight to make rubber ring press cable When wire strippi...

Page 23: ...accuracy in measurement Under normal pipe condition the Q value is in the range between 600 900 the higher the better Disadvantage of lower Q value is as follows 1 Interference of other instruments a...

Page 24: ...own Compound or Up Compound Push Down Compound or Up Compound Down key Subtract 1 from the number above cursor Up key Plus 1 to the number above cursor Compound Down To shift cursor to left Compound U...

Page 25: ...parameters in any grade of password However if users want to change parameters of converters deferent grade of parameters have to be used by the users 4 3 4 Keypad The keypad of the flow meter has 16...

Page 26: ...he value For example if the current window is on M11 and the user wants to enter 219 2345 as the pipe outer diameter then the flowing keys should be pressed For option selection windows the user shoul...

Page 27: ...up the built in logger M60 M78 windows are for setting up timekeeper and displaying software version system serial number ESN and alarms M82 window is for viewing the totalizer M86 M89 windows are us...

Page 28: ...key to enter the option selection ENT key to enter the option selection mode use keys Up and down to scroll up and down to the liner material and then press key Select No Liner if there is no liner 5...

Page 29: ...gnal quality and working status M04 Display date and time instantaneous flow rate signal strength signal quality and working status M05 Display date and time velocity signal strength signal quality an...

Page 30: ...M24 Window for selecting the transducer mounting methods Four methods can be selected V method Z method N method W method M25 Display the transducer mounting spacing or distance M26 Entry to store th...

Page 31: ...eturn 10 0AH line feeding 42 2AH 38 65535 Every set of the instrument in a network environment should have a unique IDN Please refer to the chapter for communications M47 System lock to avoid modifica...

Page 32: ...Buzzer setup If a proper input source is selected the buzzer will beep when the trigger event occurs M78 OCT Open Collector Output setup by selecting a proper triggering source the OCT circuit will c...

Page 33: ...tal number of times the flow meter has been powered on and off M 5 A scientific calculator for the convenience of field applications All the values are in single accuracy All the mathematic operators...

Page 34: ...n the calendar Initialise the calendar via window M61 Repeated reboot Hardware problem Contact Tek Trol LLC 6 6 Further Errors and Countermeasures 1 When the device indicates 0 0000 even though there...

Page 35: ...ted 1 Poor signal 2 Sensors installed improperly 3 Too much fouling or contamination 4 Liner too thick 5 Sensor cable not properly connected 1 Change measuring location 2 Clean measuring location 3 Ch...

Page 36: ...93 2 1 8 54 1 16 2 0 47 681 56 812 36 522 2 5 8 66 7 1 16 2 0 37 536 45 652 28 406 3 1 32 76 1 1 16 2 0 33 478 39 565 25 362 ADDED TOUGHNESS DURABILITY 4 5 16 108 1 16 2 5 29 420 34 493 22 319 Based o...

Page 37: ...hickness I D inch mm inch mm Min Max inch mm Min Max inch mm Min Max inch mm Min Max inch mm 1 2 15 53 64 21 20 27 3 21 50 1 40 1 70 23 32 18 25 3 4 20 1 3 64 26 60 1 1 16 26 90 1 40 1 70 15 16 23 65...

Page 38: ...07 1 7 5S 0 09 2 1 10S 0 11 3 2 STD 40 40S 0 15 3 7 XS 80 80S 0 19 4 7 160 0 30 7 5 XXS 3 4 19 1 050 26 0 07 1 7 5S 0 09 2 1 10S 0 12 2 9 STD 40 40S 0 16 3 9 XS 80 80S 0 22 3 9 160 0 31 7 8 XXS 1 25...

Page 39: ...1 5S 0 12 3 10S 0 23 5 7 STD 40 40S 0 32 8 1 XS 80 80S 0 64 16 2 XXS 4 100 4 500 114 0 09 2 1 5S 0 12 3 10S 0 24 6 STD 40 40S 0 34 8 6 XS 80 80S 0 44 12 7 120 0 54 13 5 160 0 68 17 1 XXS 5 125 5 536 1...

Page 40: ...60 0 69 17 5 80 0 85 21 4 100 1 00 25 4 XXS 120 1 13 28 6 140 14 350 14 000 355 0 16 4 5S 0 19 4 8 10S 0 25 6 4 10 0 32 7 9 20 0 38 9 5 30 0 44 11 1 40 0 50 12 7 XS 0 60 15 60 0 63 15 9 XXS 0 75 19 8...

Page 41: ...3 9 10S 0 25 6 4 10 0 38 9 5 STD 20 0 50 12 7 40 0 88 22 2 60 1 13 28 6 80 1 38 34 9 100 1 63 41 3 120 1 88 41 3 140 2 22 56 3 160 24 600 24 000 610 0 22 3 9 5S 0 25 6 4 10S 0 38 9 5 10 0 50 12 7 STD...

Page 42: ...5 230 0 51 13 9 30 236 0 56 14 9 30 236 0 60 15 10 254 0 11 10 282 0 50 13 11 10 282 0 57 14 11 40 290 0 62 16 11 40 290 0 68 17 12 304 8 13 20 335 0 54 14 13 20 335 0 62 16 13 50 343 0 68 17 13 50 34...

Page 43: ...02 51 7 1 5 Ductile Iron Pipe Standard Pipe Size Charts Nominal Pipe Size Outer Diameter Pipe Wall Thickness Class Class Class Class Class Class Class 50 51 52 53 54 55 56 in mm in mm in mm in mm in m...

Page 44: ...nealed 7 628 2 235 Copper rolled 7 448 2 270 CuNi 70 Cu 30 Ni 8 334 2 540 0 1980 5 03 CuNi 90 Cu 10 Ni 6 759 2 060 0 1579 4 01 Brass Naval 6 923 2 120 0 1744 4 43 Gold hard drawn 3 937 1 200 0 1276 3...

Page 45: ...136 40 5083 99 181 40 5097 11 48 20 4734 91 93 20 4979 33 138 20 5086 29 183 20 5095 80 50 00 4748 03 95 00 4985 89 140 00 5088 25 185 00 5094 49 51 80 4760 83 96 80 4992 45 141 80 5090 22 186 80 5093...

Page 46: ...0 98 3158 2 374 F 962 6 190 C 286 000 26 600 t Amyl Alcohol C5H12O 0 81 3 950 1 1 204 47 064 4 374 Aminobenzene 41 C6H5NO2 1 022 5 377 3 1 639 4 0 39 058 3 63 Aniline 41 C6H5NO2 1 022 5 377 3 1 639 4...

Page 47: ...52 13 55 68 F 1 26 20 C 1 Decene 27 C10H20 0 746 4 051 8 1 235 4 0 n Decene 27 C10H20 0 746 4 051 8 1 235 4 0 Diacetyl C4H6O 0 99 4 055 1 1 236 4 6 Diamylamine C10H23N 4 120 7 1 256 3 9 8 5 68 F 1 2Di...

Page 48: ...97 17 C Ethyl Bromide 46 C2H5Br 1 456 68 F 2 952 8 68 F 1 658 2 959 68 F 0 275 20 C Ethyliodide 46 C2H5l 1 950 68 F 2874 68 F 1 343 20 C 3 12 0 29 Ether C4H10O 0 713 3231 6 985 4 87 3 346 0 311 Ethyl...

Page 49: ...thylene C6H10O 0 948 4 668 6 1 423 4 0 Lithium Fluoride 42 LiF 8 152 9 1652 F 2 485 900 C 1 29 Mercury 45 Hg 13 594 4 753 9 75 F 1 449 24 C 1 226 0 114 Mesityloxide C6H16O 0 85 4 297 9 1 310 Methane 2...

Page 50: ...2 C5H10O2 4 199 36 F 1 280 2 C 4 63 n Propyl Alcohol C3H8O 0 78 68 F 4 009 2 68 F 1 222 20 C 27 427 2 549 Propylchloride 47 C3H7Cl 0 892 3 471 1 1 058 4 067 0 378 Propylene 17 18 35 C3H6 0 536 55 4 F...

Page 51: ...m CCl2F CClF2 1 563 2 571 2 32 F 783 7 0 C Triethyl Amine 33 C6H15N 0 726 3 684 4 1 123 4 47 Triethyleneglycol C6H14O4 1 123 5 275 6 1 608 3 8 1 1 1 Trifluoro 2 Chloro 2 Bromo Ethane C2HClBrF3 1 869 2...

Page 52: ...www tek trol com 50 8 Parts Descriptions Fig 20 Standard Without sensors Fig 21 Customized With sensor...

Page 53: ...s and solutions for process power and oil gas industries Tek Trol provides process measurement and control products for Flow Level Temperature Pressure Measurement Control Valves Analyzer systems We a...