10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
2-1
PREPARATION FOR USE AND INSTALLATION
SECTION 2
PREPARATION FOR USE AND INSTALLATION
2.1 Unpacking and Inspection
Upon receipt, the R1L-BI/R1L-BIR and accessories should be
carefully unpacked and removed from the shipping container.
Separate the units from the packing material and inspect both
the instrument and the accessories (if any) for any external
damage.
•
If any dents, broken, or loose parts are seen, do not use the
equipment. Notify TEGAM immediately.
•
Check that all items are present. If any items are missing,
notify TEGAM immediately.
2.2
Preparation for Use
Battery Installation
Note that the instrument is shipped with its battery removed.
Unwrap the 3 cells which comprise the battery.
Be sure the power line cord is disconnected. Check that the
Power Switch is in the OFF position. Remove the four screws
at the corners of the R1L-BI/R1L-BIR’s top (control/display)
panel. Carefully lift the unit out of its cabinet. Place the unit
upside down on a surface which will not mar or damage the
control/display panel.
WARNING
DO NOT CONNECT THE AC LINE CORD PRIOR TO REMOVING
THE UNIT FROM ITS CASE OR WHILE THE UNIT IS OUT OF
ITS CASE, OR ELSE THE MAINS VOLTAGE WILL PRESENT A
POSSIBLE SHOCK HAZARD.
CAUTION
BE SURE TO OBSERVE THE PROPER POLARITY WHEN
INSERTING THE BATTERY CELLS OR DAMAGE MAY RESULT.