Preparation for Use
2-2
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-
466
-
6100 ● FAX 440
-
466
-
6110 ●
Agency Compliance and Safety Warnings
(This section, Agency Compliance
and Safety Warnings, is controlled and may only be changed
with approval of the Notified Body.)
Ex
Hazardous Safety Information
This instrument is intrinsically safe. WARNING:
s
ubstitution of components may impair
intrinsic safety.
This instrument complies with Agency Standards
and is listed
as:
•
II 1G
•
Ex ia
IIB T4 Ga
•
Class I,
Division 1,
Group C
and
D,
T4
•
Class I,
Zone 0 AEx ia IIB T4
•
Zone 0 Ex ia IIB T4 Ga
This operation
manual contains safety information and warnings that must be observed for safe
operation under the conditions described. Failure to comply with the information and
instructions can have serious consequences. Read this manual in its entirety before using
the
instrument.
WARNING –
EXPLOSION HAZARD
To reduce the ignition risk of a flammable or
explosive atmosphere, batteries must be
changed in a location known to be non
-
hazardous.
AVERTISSEMENT – RISQUE
D’EXPLOSION
Afin de réduire le risque d’allumage dans
une atmosphère inflammable ou explosive,
les batteries doivent être remplacées dans
un endroit connu pour être non dangereux.
WARNING –
EXPLOSION HAZARD
Batteries are part of the intrinsically safe
system. Use only the specific battery type,
manufacturer, and manufacture’s part listed
in the Operation Manual. Use of other
batteries could impair intrinsic safety.
Batteries may not be mixed.
AVERTISSEMENT – RISQUE
D’EXPLOSION
Les batteries font partie du système de
sécurité intrinsèque. Utiliser uniquement le
type et le modèle spécifique de batterie et
les pièces du fabricant mentionnées dans le
manuel. L’utilisation d’autres batteries
pourrait affecter la sécurité intrinsèque.