Service Information
5
-6
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-466-
6100 ● FAX 440
-466-
6110 ● [email protected]
39.
A value will be displayed, press
to select the correct value of the 10 Ω
standard.
40.
Press
a.
“-----
” will flash periodically
41.
Short the leads of the unit as described in
Step 6
.
42.
Press the
to select 0.00.
43.
Press
a.
“-----
” will flash periodically
44.
The UUT display will then indicate as follows:
a.
“2”
b.
Line 1:
-----
c.
Line 2: 100 Ω
45.
Connect the probes to the 10
0
Ω resistance standard
46.
A value will be displayed, press
to select the correct value of the 100 Ω
standard.
47.
Press
a.
“-----
” will flash periodically
48.
“
ERROR” will begin flashing. Hold
for 3 seconds to store the alignment.
If the unit is powered off during the alignment process or
is not held for 3
seconds
at the end of the alignment process, the calibration
factors
will not be saved.