Press
a.
“-----
” will flash periodically
13.
The UUT display will then indicate as follows:
a.
“2”
b.
Line 1:
-----
c.
Line 2: 10
m
Ω
14.
Connect the probes to the 1
0 m
Ω
resistance standard
.
15.
A value will be displayed, press
to select the correct value of the 10 mΩ
standard.
16.
Press
a.
“-----
” will flash periodically
.
17.
The UUT display will indicate as follows:
a.
“1”
b.
Line 1:
-----
c.
Line 2: 100
m
Ω
18.
Short the leads of the unit as described in
Step 6.
19.
Press
to select 0.00.
20.
Press
a.
“-----
” will flash periodically
21.
The UUT display will then indicate as follows:
a.
“2”
b.
Line 1:
-----
c.
Line 2: 100
m
Ω
22.
Connect the probes to the 100
m
Ω
resistance standard
23.
A value will be displayed, press
to select the correct value of the 100 mΩ
standard.
24.
Press
a.
“-----
” will flash periodically