5.5
rAtE-P.dEt:Peak detection
Measuring rate >100Hz, the value is displayed unfiltered.
Application with logger function: Measuring of short pressure peaks or fast changing pressures
with a resolution of < 10ms.
The cyclic logger function records the arithmetic mean value, the highest and the lowest peak of
the referring time interval.
Attention: higher power consumption, measuring is sensitive to noise (EMI,..).
5.6
rAtE-FASt: Fast: fast filtered measuring
Measuring rate >100Hz, the value is filtered slightly (higher noise immunity than P.dEt, small
peaks will be filtered out), apart from that identical behaviour like P.dEt.
5.7
(P.oFF) Power off time
If there won’t be pressed any key and no interface communication takes place for the time of the
power off time setting (P.Off), the device will be switched off automatically to save battery power.
If P.oFF = oFF then the automatic switch off is deactivated.
5.8
(Adr.) Interface address
In preparation for a 5 channel-interface converter
5.9
(AL.) Alarm
There are three possible settings: Alarm off (AL. oFF), on with horn sound (AL. on), on without
horn sound (AL. no.So).
Following conditions will display an alarm, when the function is activated (on or no.So):
-
Value is below lower (AL. Lo) or above upper alarm rail (AL.Hi).
-
Sensor error (Sens Erro)
-
Low battery (bAt)
-
Fe 7: System error (always with sound)
In case of an alarm, the alarm arrow flashes, and when polling the interface the prio-flag is set in
the returned interface message.
5.10 (CLOC) Real time clock
The real time clock is used for the logger function: Recorded values are also containing the point
of time, when they were measured. Please check the settings when necessary.
If the battery was replaced the referring menu ‚CLOC‘ will automatically be started.