Betriebsanleitung/Operating manual
E3905 / P3394
30
BD_BE 763 b
www.tecsis.de
Button functions:
Max-Function:
Pressing ´max´ shows the maximum of the measured values. Pressing it
again hides it. To clear the max memory press key ´max´ for >2 seconds.
Min-Function:
Pressing ´min´ shows the minimum of the measured values. Pressing it
again hides it. To clear the min memory press key ´min´ for >2 seconds.
Hold-Function*:
By pressing ´Store/Quit´ the last measuring value will be held in the
secondary display. Pressing it again hides it.
(*only when main menu logger = ‚off‘ is chosen).
Logger-Functions:
Activation with key ´Store/Quit´, if in the main menu a logger function
was chosen before (see chapter 6).
Tara-Function:
By pressing ´Tara´ the display will be set to 0. All measurings from then on
will be displayed relatively to the set tare value. When tara function is
activated, the arrow "Tara" appears in the display.
To deactivate tare function press ´Tara´ for >2 seconds
Please note: Activating / deactivating tara clears
the max- & min-memories.
Zero point adjust:
(for gauge pressure sensors only) If there is no pressure applied to the
pressure ports the device will display 0. If there is a permanent deviation
(and device is operated under steady conditions), a permanent zero point
adjustment can be carried out.
To carry out the adjustment press ´Tara´ for approx. 5 seconds.
(Please note: A zero-point adjustment can only be carried out if the
difference between the value on display and the value calibrated on site is
less than 2%! E.g. for the measuring range of 0 ..+25.00mbar,
=>zero-point adjustment up to 0.50mbar possible)
To recall the manufacturers’ calibration press ´Tara´ for approx. 7 seconds.
Please note: If a zero-point adjustment was carried out, this will be signalled
by the short displaying of „NuLL Corr“ when switching on
the device.
Differential press**:
Pressing ´set/Menu´ shows the difference of channel 1- channel 2 (CH1-
CH2) in the secondary display. Pressing it again will undo it (**only when
2 channel version and 2 sensors connected).