Accuraspray4.0
Pin Name
Description
Logic
1
24 VDC IN Input/Output reference Voltage
+
2
0 VDC IN
Input/Output reference Voltage -
3
IN 1
Start measurement sequence
(the data acquired by the
Accuraspray will be saved) when
logic is 1 and stop when logic is
0
Logic 1 : 11-24V
above 0VDC IN
Logic 0 : 0-5V
above 0VDC IN
4
OUT 1
Wait for data acquisition.
This output indicates that the
sensor is acquiring its initial data
buffer.
Once the buffer is filled, this
output will fall.
Logic rising edge:
Acquiring
initial
buffer
Logic falling edge:
Initial
buffer
acquired
5
OUT 2
Measurement valid.
Following the acquisition of the
first buffer this output 1 will be
raised when measurements are
valid.
If the measurements are invalid,
error messages
will be displayed
in the user interface and
measurements results should be
ignored.
Logic 1 : 24 VDC IN
Logic 0 : 0 VDC IN
6
OUT 3
All measurements are within the
green zone
.
Logic 1 : 24 VDC IN
Logic 0 : 0 VDC IN
7
OUT 4
At least one measurement is
within the
yellow zone
.
Logic 1 : 24 VDC IN
Logic 0 : 0 VDC IN
8
OUT 5
At least one measurement is
within the
red zone
.
Logic 1 : 24 VDC IN
Logic 0 : 0 VDC IN