Commissioning
Proline Promass F 500 EtherNet/IP
150
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Switch-on value
• In the
Operating mode
parameter, the
Switch
option is selected.
• In the
Switch output
function
parameter, the
Limit
option is selected.
Enter measured value for the
switch-on point.
Signed floating-point
number
Country-specific:
• 0 kg/h
• 0 lb/min
Switch-off value
• In the
Operating mode
parameter, the
Switch
option is selected.
• In the
Switch output
function
parameter, the
Limit
option is selected.
Enter measured value for the
switch-off point.
Signed floating-point
number
Country-specific:
• 0 kg/h
• 0 lb/min
Switch-on delay
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Limit
option is selected
in the
Switch output
function
parameter.
Define delay for the switch-on
of status output.
0.0 to 100.0 s
0.0 s
Switch-off delay
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Limit
option is selected
in the
Switch output
function
parameter.
Define delay for the switch-off
of status output.
0.0 to 100.0 s
0.0 s
Failure mode
–
Define output behavior in
alarm condition.
• Actual status
• Open
• Closed
Open
Invert output signal
–
Invert the output signal.
• No
• Yes
No
*
Visibility depends on order options or device settings
10.5.10 Configuring the relay output
The
Relay output
wizard guides the user systematically through all the parameters that
have to be set for configuring the relay output.
Navigation
"Setup" menu → Relay output 1 to n
‣
RelaisOutput 1 to n
Switch output function
Assign flow direction check
Assign limit
Assign diagnostic behavior