CARDIATEK UG 202001-R01
33
Maintenance Guide
CLEANING THE CARDIATEK SYSTEM
CLEANING PROCESS
The Cardiatek Bracket, Wall Mount and Floor Mount can be cleaned by using a pressure washer. The mounting bracket
and system may show some signs of oxidation or discoloration from continuous washing, however, degradation of the
cot/stretcher’s performance or functionality will not be affected due to power washing if cleaning instructions are
properly followed.
The material used to manufacture the Cardiatek System is high-grade aluminum, it is very durable and resistant to
corrosion, however high levels of sodium or a high concentration of sodium or acid may prematurely corrode the
product. Always rinse with soft soap and water if exposed.
9
Use of a power washer can accelerate removal of contaminants collected during the use of the product.
9
Rinse with clean water.
9
Towel dry all components.
9
Allow all other components to air dry.
9
Avoid over saturation and ensure that the product does not stay wet longer than the cleaner manufacturer’s
guidelines for proper disinfecting.
WARNING
9
When cleaning, always use appropriate personal protection equipment (PPE) based on established protocols (e.g.,
gloves, eyewear, etc.).
CAUTION
9
Do not steam clean or use ultrasonic cleaners on the system or any of its components.
9
Do not immerse the metal parts/components in water.
9
Maximum water temperature should not exceed 180°F/82°C
9
Maximum water pressure should not exceed 1500 psi/103.5 BAR . If a pressure washer is being used to clean the
unit, the pressure nozzle must be kept a minimum of 24 inches (61 cm) from the unit.
CLEANING SOLUTIONS
Phenolic type or quaternary type disinfectants can be used (excluding Virex® TB). Iodophor type disinfectants,
Peracetic Acid, Chlorine, and especially Peroxide Oxygen are not recommended for use because staining and corrosion
may result.
Suggested cleaners for the Cardiatek System (Bracket, Wall Mount and Floor Mount):
9
Quaternary Cleaners (active ingredient - ammonium chloride)
9
Phenolic Cleaners (active ingredient - o-phenyl phenol)
9
Chlorinated Bleach Solution (3.25% - less than 1-part bleach to 100 parts water)
Summary of Contents for Cardiatek
Page 1: ...CARDIATEK USER GUIDE ...
Page 43: ...www technimount com CARDIATEK UG 202001 R01 43 Return To Table of Contents SAE Certification ...
Page 44: ...Return To Table of Contents 44 CARDIATEK UG 202001 R01 www technimount com FAA Certification ...
Page 45: ...www technimount com CARDIATEK UG 202001 R01 45 Return To Table of Contents CE Certification ...