background image

Sample

Concentrator

operator'S manual

Issue 14

10/15

Summary of Contents for FSC400D

Page 1: ...Sample Concentrator operator s manual Issue 14 10 15...

Page 2: ......

Page 3: ...fety 3 Specification 3 Installation 3 Diagram of Sample Concentrator 4 Guarantee 5 operation 6 Preparation 7 Operation 8 ADDITIONAL INFORMATION 9 User maintenance 9 Solvent evaporation temperatures 10...

Page 4: ...rected over the surface of the samples displaces the evaporated solvent from above the liquid s surface This significantly increases the rate of sample concentration The unit provides accurate and rep...

Page 5: ...to suit Techne Dri Blocks Gas Intake Nozzle Diameter 6 35 mm Height 530 mm Width 240 mm Depth 295 mm Weight 3 5 Kg Installation For packing purposes the Sample Concentrator is supplied unassembled in...

Page 6: ...xcess of 2 psi will cause gas to be vented and wasted 4 Place the Techne Dri Block heater onto the base of the Sample Concentrator It is positioned by means of two spigots on the base board which alig...

Page 7: ...under this guarantee should be obtained from the supplier Notwithstanding the description and specification s of the units contained in the user manual Techne hereby reserves the right to make such ch...

Page 8: ...test tubes to be placed in the heater The needle guide holes are labelled with a letter corresponding with the type of insert block used in the heater These are listed in the following table As an ex...

Page 9: ...ion Build up the desired pattern of needles to suit the arrangement of test tubes to be used Do not put a needle in any position which will not have a corresponding sample to be concentrated as this w...

Page 10: ...required height above the liquid surface 3 Set the heater to the required temperature and adjust the gas flow to the desired rate Remember not to exceed 2 psi 4 To increase the rate of concentration t...

Page 11: ...ad After many repeated needle insertions the self sealing property of the sealing pad may start to degrade Spare pads can be obtained from your supplier or from Techne Replacing the pad is simply acco...

Page 12: ...35 Atmospheric 1 2 dichloroethane Ethylene chloride 83 210 1 2 dichloroethene Cis acetylene dichloride 60 479 1 2 dichloroethene Trans acetylene dichloride 48 751 2 propan 2 yloxypropane Isopropyl eth...

Page 13: ...ethane Carbon tetrachloride 76 300 1 1 2 2 tetrachloroethene Tetrachloroethylene 121 53 Oxolane Tetrahydrofuran THF 67 357 Toluene Methylbenzene 111 77 1 1 2 trichloroethene Trichloroethylene 87 183 O...

Page 14: ...B B B B B B B B B B B B B B B A A A A A A A A A A A A A A A A A A A A A A A A A A D D D D D D A A A A A A A A A A A A A A A C C C C C C C C C C C C C C C C C F E E E E E E F F F F F B B B B B B B B B...

Page 15: ...7210 Pack of 100 x 127mm long needles FSC4NCS Pack of PTFE coated needles 100 x 76mm long FSC4NCL Pack of PTFE coated needles 100 x 127mm long 6101608 Spigot 6101604 Sealing pad 6101605 Support strip...

Page 16: ...l contact bibby scientific com Web www bibby scientific fr Bibby Scientific US 3 Terri Lane Suite 10 Burlington NJ 08016 USA Tel 1 609 589 2560 Fax 1 609 589 2571 Email labproducts techneusa com Web w...