background image

FLOMAT FX 

Electromagnetic

 

insertion sensor 

Series 

Instructions Manual 

R-MI-FlomatFX   Rev.: 1 English version 

instrumentation

for fluids 

Summary of Contents for FLOMAT FX FlomatFX

Page 1: ...FLOMAT FX Electromagnetic insertion sensor Series Instructions Manual R MI FlomatFX Rev 1 English version instrumentation for fluids...

Page 2: ...4 MOUNTING THE INSERTION FITTING 7 5 MOUNTING THE SENSOR 9 5 1 Tightening torque 9 5 2 Electronic converter connection 9 6 CONFIGURATION 9 6 1 Fc factor 9 6 2 Pipe diameter 9 7 POSSIBLE PROBLEMS WHEN...

Page 3: ...k the instrument removing any packaging material that could be attached to the sensor Do not remove any grease from the collar that fits the sensor to the electronics housing 2 2 Storage temperature 2...

Page 4: ...sensor must be BS 1042 2 2 1983 standard 3 1 Sensor position The most adequate position is in the side of the pipe In this way deposits of particles on the electrodes and air pockets at the top of the...

Page 5: ...different conductivities are mixed it is necessary to install the sensor a minimum of 50 x DN from the point of mixture in order to obtain a uniform conductivity of the liquid and stabilize the readi...

Page 6: ...G in the range of 20 150 Hz according to IEC 068 2 34 3 7 Magnetic fields Strong magnetic fields close to the sensor should be avoided 3 8 Temperature In open air installations it is recommended to i...

Page 7: ...e thickness Peel off the bottom part of the label Once the insert is placed into its final position where the label was cut must coincide with the exterior of the pipe This ensures that the measuring...

Page 8: ...us example if the gasket was 5 mm thick the distance that the fitting should extend above the exterior wall would be H 55 3 5 53 mm DN Sensor length Fitting length C H 80 101 93 10 88 s 100 12 5 85 5...

Page 9: ...In the event that during the installation of the sensor the electronics or the sensor cable were disconnected simply replace the two connectors on the sensor push the mesh above the head tight the si...

Page 10: ...7 2 1 Integration time It is the time during which the average value is calculated In the supposition that the instrument takes 10 readings per second if an integration time of 5 seconds is selected...

Page 11: ...housing EN 1 4404 AISI 316L PVDF Electrodes head PVDF Gasket NBR 9 2 Insert fitting connection G2 1 4 thread DIN flange special flange 9 3 General characteristics Accuracy 3 5 with respect to measure...

Page 12: ...due to misuse being excluded all responsibility due to any other damage or the effects of wear caused by the normal use of the devices Any consignment of devices for repair must observe a procedure w...