3.
Installation
2011-09
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
23
3.2
Plate Carrier Transport Lock
STOP
Caution
Before the instrument is switched on for the first time,
it should be left to stand for at least 3 hours, so there is no
possibility of condensation causing a short circuit.
STOP
Caution
Remove the transport lock before operating the instrument.
The instrument is delivered with the plate carrier locked into place, so that it
cannot be damaged. Before the instrument can be used, the transport locks must
be removed using the following procedure:
1. Switch ON the computer and install the corresponding software on the
computer (i-control, Magellan or EVOware).
2. Ensure that the computer is switched OFF and the instrument's main power
switch on the back panel of the instrument is in the OFF position.
3. Connect the computer to the instrument only with the delivered USB interface
cable.
4. Insert the power cable into the main power socket (with protective earth
connection) on the back panel of the instrument.
All connected devices must be approved and listed as per IEC 60950-1
Information Technology Equipment – Safety or equivalent local standards.
5. Open the plate door manually and loosen the two outer screws from the
Transport Lock (2.5 mm Allen key is supplied).
6. Switch ON the instrument using the main power switch on the back panel of the
instrument.
7. Switch ON the computer and start the corresponding software on the computer
(i-control, Magellan or EVOware).
8. Connect the INFINITE M1000 PRO instrument via the software.
9. The software displays a message stating that the instrument is parked and
requests the loosening of the two outer screws from the Transport Lock -
confirm with
OK
.
Figure 3