3.
Installation
2008-07
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
49
3.2 Removal of the Transport Locks
STOP
Caution
Remove the transport lock before operating the instrument.
The instrument is delivered with the plate carrier and filter carrier/cuvette carrier
locked into place, so that they cannot be damaged. Before the instrument can be
used, the transport locks must be removed using the following procedure:
1. Ensure that the instrument is disconnected from the main power supply.
2. Open the plate carrier door and filter carrier/cuvette carrier door.
3. Remove the screw from the transport lock, which extends into the filter
carrier/cuvette carrier compartment from the plate carrier compartment
(left transport lock in picture above).