6. Instrument Features
138
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
6.3
Fluorescence Intensity and
Time Resolved Fluorescence (TRF)
Parameters
Characteristics
Wavelength Range -
Infinite M200 PRO:
230 - 850 nm
selectable in 1 nm steps
Wavelength Range -
Infinite F200 PRO:
230 - 850 nm
Standard Filter:
Not applicable –
instruments are equipped with user-defined filters
Parameters
Instrument type
Characteristics
Wavelength Range
standard
Excitation: 230 – 600 nm
spectrally-
enhanced
Excitation: 230 – 850 nm
standard
Emission: 330 – 600 nm
spectrally-
enhanced
Emission: 280 – 850 nm
Gain Setting
Values
Measurement Range
Manual
1 - 255
0 - 60,000 RFU
Optimal
automatic
0 - 60,000 RFU
Calculated from
Well
automatic
0 - 60,000 RFU
TRF Parameters
Characteristics
Integration Time
10 - 2000 µs
Lag Time
0 - 2000 µs