5. Operating the Instrument
114
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
5.11.2 Washing
Before the instrument is switched off, it is recommended to perform a wash
procedure to clean the injector system.
Washing can be performed by using the i-control software or by using the
hardware buttons on the injector box.
Before performing the washing procedure:
1.
Fill the storage bottles with the appropriate wash reagents (distilled water,
70 % ethanol, etc…) and insert the feeding tubes. Make sure, that the tubes
reach the bottom of the bottle.
2.
Remove the injector from the carrier slot and insert it into the service position
of the injector box.
3.
Put an empty container under the injector.
Washing Procedure (i-control):
Adjust the parameters on the
Wash
tab of the
Injector Maintenance
dialog box
in the
Settings
menu.
1.
Start the washing procedure by clicking the Start wash.
2.
Select one of the injectors
Injector A
or
Injector B
or
Injector A and B
.
3.
Select the
Piston Strokes
(1 – 60; 1 stroke equals 1 ml)
4.
Select the
Wash Speed
(100 - 300 µl/sec).
5.
Select the
Refill Speed
(100 – 300 µl/sec.) or select
Refill Speed equal to
Wash Speed
.
6.
Click
Start wash
to start the wash procedure.
7.
Click
Close
to exit the dialog box.