4 - 2
Balance Kit Operating Manual, 393278, en, V1.0
4 - Putting into Operation
Setting up the Balance Kit and the SAG 285/01
4.1.2 Unpacking and Inspection
1
Unpack the Balance Kit and the SAG 285/01.
2
Ensure that no parts are missing or damaged.
3
Check whether the voltage range of the SAG 285/01 supply unit corresponds
with the local mains voltage.
4
Contact the supplier of the SAG/01 or your nearest Tecan representative if
any parts are missing or damaged.
4.1.3 For Your Safety
WARNING
Damage to the system and danger of electric shock.
Always switch the instrument and the computer off before making any electric
connections or opening the instrument’s access door.
4.1.4 Preparing the Weighing Cell
4
Carefully place the Weighing Cell onto the adapter plate with the feet inserted
into the corresponding positioning holes.
5
Check whether the Weighing Cell is seated properly on the adapter plate and
does not wobble. If necessary adjust the cell by turning the feet until the cell is
level and stable.
6
Carefully tighten the counternuts to secure the feet when done.
7
Place the centering ring (D) on the cell and insert the cone of the weighing pan
(E) into the corresponding hole as shown in the figure.
To prepare the Weighing Cell:
1
Screw the counternuts on the
screw bolts of the feet (F)
down until the stop.
2
Screw the feet (F) into the
corresponding mounting
holes at the bottom of the
Weighing Cell (H).
3
Place the adapter plate (G)
on a the worktable in the grid
position where you want to
perform the Gravimentric test
later.
D
E
F
G
H