3 - 8
Balance Kit Operating Manual, 393278, en, V1.0
3 - Technical Data
Requirements
3.3.3 Access Rights, Passwords
To carry out the Balance Setup function and the Gravimetric Test with the
Instrument Software you need the following access rights:
Version V4.6x or earlier: No password required.
Version V4.7 or later: From version V4.7 onwards, all users need a user
account, together with a user name and a password.
–
There are several predefined user groups with different authorization
levels.
–
The Balance Setup and the Gravimetric Test can be carried out by users
with SnS_Customer access rights.
Note:
For detailed information about user accounts, passwords and access rights
refer to the Instrument Software Manual (for software version V4.7 or later),
section “Tecan User Administration System”.
3.3.4 Syringes and Liquid for Gravimetric Test
Syringes: 250
µ
l, 500
µ
l, 1 ml.
System liquid: Demineralized or de-ionized and degassed water. Water
temperature should correspond to ambient temperature.
Pipetting liquid: Tap water. Water temperature should correspond to ambient
temperature.