Teac TU-MXR2(T)-G3 Instructions For Use Download Page 1

D01399001A

Z

Strain gauge load cell Instructions for Use

TU-MXR2(T)-G3

Tension/Compression Load Cell

Introduction

Thank you for purchasing the TU-MXR2(T)-G3 load cell.
Please read this document completely before using this 
load cell to achieve its best performance and ensure safe 
and proper operation.

 

o

Company names and product names in this document are the 
trademarks or registered trademarks of their respective owners.

Included accessories

If anything is missing or damaged, contact the retailer 
where you purchased the product.

Test report × 1
Jig for tightening × 1
Instructions for Use (this document) × 1

IMPORTANT SAFETY INSTRUCTIONS

V

WARNING

If something abnormal occurs

Request repair from the retailer where you purchased 
the product.

Do not open the cover.

Never remove the cover from this unit. Doing so could 

cause malfunction. Request inspection and repair from the 
retailer where you purchased the product. Do not alter this 
unit. Doing so could cause malfunction.

Do not put foreign objects or water, for example, 
into the unit.

Do not place a container that holds water, for example, on 
top of this unit. If liquid is spilled, for example, and enters 
the unit, this could cause malfunction.

Do not use the unit with any power supply voltage 
other than that specified.

Do not use the unit with any power supply voltage other 
than that specified. Doing so could cause malfunction.

V

CAUTION

Unsuitable installation locations

Do not place the unit in the following types of locations. 
Doing so could cause malfunction.

 

o

Locations where it might be exposed to smoke or steam, 
such as near a kitchen table or humidifier

 

o

Unstable locations, including unsteady stands and 
tilted places

 

o

Locations that are very humid or dusty

 

o

Locations that are exposed to direct sunlight

When not using the unit for a long time

For safety, cut the power supply when not using this unit 
for a long time.

Do not operate a damaged unit.

Precautions for use

 

o

This unit is not built to be water or splash resistant, 
and it cannot be used in conditions when the relative 
humidity is high. Moreover, use in atmospheres with 
corrosive gases should be avoided.

 

o

Be careful to prevent water, oil and other substances 
from getting on the unit.

 

o

Avoid use in conditions where condensation could occur.

 

o

Connect cores to the load cell after discharging (elim-
inating) static electricity from your body.

 

o

If the surrounding temperature changes suddenly, the 
values output by this device could become unstable, 
making accurate measurement impossible.
(This could occur, for example, in a location blown by 
warm or cold air.)

 

o

Conduct load calibrations periodically.

Installation procedures

Insert a screw into the screw hole in the center part of the 
load cell, and fix the screw to a flat attachment surface. 
(Output precision could be affected if the attachment 
surface is not flat.) The size of the screws used depends on 
the rated capacity (model name) of the load cell as shown 
in “Recommended tightening torques when attaching to 
load jigs and surfaces” belo
w.

Recommended tightening torques when attaching 
to load jigs and surfaces

Rated 

capacity

Dimension 

of load 

cell screw 

hole

Effective 

screw 

depth

Body 

material

Tightening 

torque 

tolerance

10 N

M3

3.5 mm

Aluminum

0.3 N·m

20 N

0.7 N·m

50 N

1.1 N·m

100 N

1.5 N·m

200 N

Steel

6.0 N·m

500 N

M4

4.5 mm

6.0 N·m

Use screws with a strength grade of 12.9.
In addition, see the “Effective screw depth” in the dimensional 
drawing for the length of thread engagement of the screw 
and the mounting part (internal threads).
Please consult with us if you are uncertain about screw 
selection, for example.

ATTENTION

 

o

When attaching a load button or rod end, for example, 
to the screw/bolt area on the top of the load cell, use 
the included jig to hold the area shown with diagonal 
lines at the top and tighten the screw or nut.

Jig for tightening

Wrench  

(commercially-available)

 

o

In the same way, when attaching a rod end or base, 
for example, to the screw/bolt area on the bottom 
of the load cell, use the included jig to hold the area 
shown with diagonal lines at the bottom and tighten 
the screw or nut.

Jig for tightening

Wrench  

(commercially- 

available)

 

o

When tightening the screw or nut attached to the load 
cell, holding the screw/bolt part on the opposite end 
(shown with diagonal lines), will cause twisting force 
to be applied to the gauge and break the load cell.

Jig for tightening

Wrench 

(commercially-available)

 

o

When removing it, follow the attachment procedures 
in reverse order. On the top or the bottom side of the 
load cell that has 14mm width across flats, the side that 
has the nut, for example, to be loosened, use the jig 
for tightening to remove it so that force is not applied 
to the load cell. 

 

o

Attach a load button, for example, when using with 
compression.
Tighten with the torque shown in the table above 
to prevent the load button from becoming loose. 
When removing an attached load button, be sure not 
to apply torque that exceeds the tightening torque 
shown in the table.
A thread-locking fluid can be used on the screw, but 
be aware that this could prevent removal of the load 
button. Please prepare a load button to attach or 
consult with us.

 

o

When using with a tensile load, attach a rod end or 
tensile load adapter.

 

o

When attaching a rod end to a load cell, use a nut to 
prevent loosening. When tightening this nut for pre-
venting loosening, secure the rod end. Do not tighten 
the nut by securing the load cell. If twisting force is 
applied between it and the central screw part, the load 
cell could be damaged.

Hold the rod 

end in place

Nut

Do not hold 

the load cell 

in place.

 

o

Always use screws/bolts that are suitable for the rated 

capacity when attaching a base plate or tensile load 

adapter, for example, to the load cell. Moreover, tighten 
all of the attachment screws/bolts evenly.

Reviews: