Strain gauge load cell Instructions for Use

TT-FR(T)-G6

Tension type Load Cell

Introduction

Thank you for purchasing the TT-FR(T)-G6 load cell.
Please read this document completely before using this 
load cell to achieve its best performance and ensure safe 
and proper operation.

Included accessories

If anything is missing or damaged, contact the retailer 
where you purchased the product.

Test report × 1
Instructions for Use (this document) × 1

Company names and product names in this document are the trademarks 
or registered trademarks of their respective owners.

IMPORTANT SAFETY INSTRUCTIONS

V

WARNING

If something abnormal occurs

Request repair from the retailer where you purchased 
the product.

Do not open the cover.

Never remove the cover from this unit. Doing so could 
cause malfunction. Request inspection and repair from the 
retailer where you purchased the product. Do not alter this 
unit. Doing so could cause malfunction.

Do not put foreign objects or water, for example, 
into the unit.

Do not place a container that holds water, for example, on 
top of this unit. If liquid is spilled, for example, and enters 
the unit, this could cause malfunction.

Do not use the unit with any power supply voltage 
other than that specified.

Do not use the unit with any power supply voltage other 
than that specified. Doing so could cause malfunction.

V

CAUTION

Unsuitable installation locations

Do not place the unit in the following types of locations. 
Doing so could cause malfunction.

 

o

Locations where it might be exposed to smoke or steam, 
such as near a kitchen table or humidifier

 

o

Unstable locations, including unsteady stands and 
tilted places

 

o

Locations that are very humid or dusty

 

o

Locations that are exposed to direct sunlight

When not using the unit for a long time

For safety, cut the power supply when not using this unit 
for a long time.

Do not operate a damaged unit.

Precautions for use

 

o

This unit is not built to be water or splash resistant, 
and it cannot be used in conditions when the relative 
humidity is high. Moreover, use in atmospheres with 
corrosive gases should be avoided.

 

o

Be careful to prevent water, oil and other substances 
from getting on the unit.

 

o

Avoid use in conditions where condensation could occur.

 

o

Connect cores to the load cell after discharging (elim-
inating) static electricity from your body.

 

o

If the surrounding temperature changes suddenly, the 
values output by this device could become unstable, 
making accurate measurement impossible.
(This could occur, for example, in a location blown by 
warm or cold air.)

 

o

If a cable of this unit needs to be bent and shifted, make 
the curvature of the bent part at least 60 mm. Do not 
apply tension to the cable.

 

o

Conduct load calibrations periodically.

Installation procedures

Ø34

Ø12H7

16

 0

-0.13

13.3

Ø51

Rod end bearing

TF-RR112F

Load cell

TT-FT(T)

Set screw

Rod end bearing

TF-RR112F

Gap

(2 mm)

 

o

The top and bottom surfaces each have a single male 
screw for attachment. Attach rod end bearings to 
them for use.

 

o

Lock nuts, for example, cannot be used to tighten the 
rod end bearings to the load cell. The load cell attach-
ment screw parts are chamfered, though, so use these 
surfaces to secure the set screws.

 

o

After connecting rod end bearings, tighten the set 
screws to prevent rotation.
During inspection before use, confirm that the set screws 
are not loose and that there are no signs of turning.

 

o

After installing a load cell, confirm that the cable will 
not be pulled.
Pulling on the cable could result in error.

 

o

Verify that safety measures to prevent falling and drop-
ping have been implemented.

Electrical connection of load cell with 

built-in TEDS

 

o

Connect as shown in the illustration below. Incorrect 
connections could result in inability to balance and 
in errors occurring in the output voltage when loads 
are applied.

Using a cable with bare lead wires

Red 

Input (+)

Black 

Output (−)

Blue 

Input (−)

White  Output (+)

Orange  TEDS signal (+)

Green  TEDS signal (COM)

Yellow Shield

TEDS

 

o

This unit has a built-in TEDS function.
The orange and green cores in the cable and the F and 
G pins in the connector are wired for TEDS.

 

o

This unit does not support remote sense.

 

o

See the operation manuals of indicators and strain 
amps that support remote sense for how to connect 
sensors with those units.

 

o

The shield is not connected to the main body of this 
product. For this reason, if grounding is necessary 
because of external noise or another issue, arrange to 
ground the shield to a part other than the body of this 
unit, for example.

 

o

Since the cable is directly connected to this unit, use 
a specialized cable to increase the length. (Please 
consult with us.)

 

o

When conducting insulation resistance tests, limit them 
to the red, black, blue and white cores. Do not apply to 
the TEDS cores (orange and green).

Using a connector (optional)

A  Input (+)

B  Output (−)

C  Input (−)

D  Output (+)

F  TEDS signal (+)

G  TEDS signal (COM)

E Shield

NDIS-compatible connector

TEDS