6-5-5. Setting value loading command format
Command numbers 0000 to 5802 in “6-6. Commands” on page 27 are transmitted in this format.
Data is fixed to 6 characters. If the data lacks 6 characters, each opening will be filled with a “0”.
Command
Hold Mode Setting (4001) example
HEX 0x23 0×30 0×31 0×34 0×30 0×30 0×31 0x45 0x30 0×0D
ASCII
#
0
1
4
0
0
1
E
0
CR
ID number
Command No.
Checksum
Response
Hold Mode Setting (4001) example
HEX 0x06 0×30 0×31 0×34 0×30 0×30 0×31 0×30 0×30 0×30 0×30 0×30
0×0D 0×0A
ASCII ACK
0
1
4
0
0
1
0
0
0
0
0
×
×
×
CR
LF
ID number
Command No.
6 data characters
Checksum
6-5-6. TEDS command format
Command numbers 6001 to 6007 in “6-6. Commands” on page 27 are transmitted in this format.
Data is fixed to 8 characters. If the data lacks 8 characters, each opening will be filled with a “0”.
Command
Example of TEDS maximum rated capacity (6002)
HEX 0x23 0×30 0×31 0×36 0×30 0×30 0×32 0×32 0×39 0×0D
ASCII
#
0
1
6
0
0
2
2
9
CR
ID number
Command No.
Checksum
Response
Example of TEDS maximum rated capacity (6002)
HEX 0x06 0×30 0×31 0×36 0×30 0×30 0×32 0×32 0×30 0×30 0×32 0×30 0×30 0×30 0×30 0×41 0×44 0×0D 0×0A
ASCII ACK
0
1
6
0
0
2
2
0
0
2
0
0
0
0
A
D
CR
LF
ID number
Command No.
8 data characters
Checksum
Data
The decimal point position is 2, so the rated capacity is 200.00.
See “6-5-8. Real-time value/hold value format” on page 26 for details about the data format for TEDS maximum rated capacity (6002).