41
6. Condition Setting
6-4. Digital Zero Offset
The set value is subtracted from the measured value (digital
zero).
6-5. Digital Zero
6-5-1. Enable Digital Zero
When Enable Digital Zero is set to OFF, the Digital Zero func-
tion will not be executed when you press and hold the ZERO
button.
6-5-2. Digital Zero Limit Value
Set the range for Digital Zero capture. (The setting value unit is
the same as for the indicator value.)
ATTENTION
If the current sensor input value exceeds the Digital Zero Limit
setting value, “Digital Zero Limit error” appears and the indica-
tor value does not become zero.
Even when Enable Digital Zero is set to ON, if Cal. Value Lock is
set to OFF, the zero function will not be executed even when
the ZERO button is pressed and held.
NOTE
If you execute the Clear Digital Zero function, the current sen-
sor input value is shown with the Zero Balancing value as zero.
6-5-3. Clear Digital Zero
The display adjustment that has been made with Digital Zero
is reverted so that the unadjusted value is shown again.
When Zero Balancing has been conducted, the value is shown
as zero.