37
5-8. Select EU
Select the indicator value unit.
Select the unit that corresponds to the indicator value.
Options: N, kN, kPa, MPa, g, kg, sht, ton, mN·m, N·m, kN·m, none
NOTE
The display unit is shown next to the indicator value, but it has
no effect on internal calculations.
For example, the calibration value will not change even if the
display unit is changed from “N” to “kN”.
5-9. Sensor input logic
The sensor input logic can be reversed artificially.
Normally, “Standard” should be used.
NOTE
“Reversed” does not electrically reverse the input.
5. Calibration
5-10. Zero Balancing
You can conduct Zero Balancing calibration even without
resetting the calibration value.
1
Set Cal. Value Lock to OFF.
See “5-1-1 Locking and unlocking calibration values” on
page 31.
2
With no load on the sensor, press and hold the
ZERO button.
During zero balancing, a pop-up appears to show that it
is in progress.
If a calibration error appears, conduct countermeasures
according to the error message.
3
Set Cal. Value Lock to ON.
5-11. Digital Zero
The current indicator value becomes zero.
When Cal. Value Lock is set to ON, press and hold the ZERO
button.
o
The range for Digital Zero is the range set by Digital Zero
Limit.
o
Digital Zero is cleared by putting the unit into standby or
using Clear Digital Zero.
ATTENTION
o
This only functions when Cal. Value Lock is ON and Enable
Digital Zero is ON.
o
When Cal. Value Lock is OFF, this function is replaced by
Zero Balancing.