PERFORMANCE SPECIFICATION
5.1.
Insulation resistance at 1000V DC:
Not less than 10 megohms between mains input to frame
5.2.
Applied voltage test
:
2.0kV rms for 1 minute between mains input and frame
5.3.
Induced voltage test:
± 35kV DC for 5 minutes
5.4.
Flashover test:
6 flashovers at 30kV DC on each output direct to earth
5.5.
Normal voltage test:
± 30kV DC on no load for 30 minutes
5.6.
Accuracy of instruments:
Output kV meters ± 1.5% at full scale
Output mA meters ± 2.5% at full scale on each range