71
Key Pump Featur
es
2
Chapter 5 – Infusion Site Care and Loading Cartridge
3.
Hold the pump vertically to
ensure any air in the cartridge will
be dispelled first. Tap
START
.
The pump will beep and vibrate
regularly while the tubing is being
filled.
Z
Z
“Starting Fill” screen is displayed.
The following are approximate
amounts of insulin to fill different
tubing lengths:
»
15–20 units for 23 inch tubing
»
20–25 units for 32 inch tubing
»
25–30 units for 42 inch tubing
4.
Tap
STOP
after 3 drops of insulin
are seen at the end of the infusion
set tubing.
Z
Z
“Stopping Fill” screen is displayed.
Z
Z
“Detecting Insulin” screen
is displayed.
5.
Verify that drops are seen and
tap
DONE
.
•
If you do not see drops, tap
FILL
.
The Fill Tubing screen appears,
repeat steps 3 to 5 until you see 3
drops of insulin at the end of the
tubing.
•
If you did not tap STOP, the “Max
fill amount reached!” screen will
appear. Do one of the following:
a.
If you are finished filling the tub-
ing, tap
DONE
.
b.
If you want to fill the tubing with
more than 30 units, tap
FILL
to
go back to the Fill Tubing screen.
6.
Fill Tubing is complete.
NOTE: Initial Display of
Insulin Level
After tubing fill is complete, when the pump
returns to the Home Screen, an estimate
of how much insulin is in the cartridge is
displayed in the upper right portion of the
screen. You will see one of the following on
the screen:
»
+ 40 u More than 40 units
detected in the cartridge
»
+ 60 u More than 60 units
detected in the cartridge
»
+ 120 u More than 120 units
detected in the cartridge
»
+ 180 u More than 180 units
detected in the cartridge
»
+ 240 u More than 240 units
detected in the cartridge
After 10 units are delivered, an actual
number of units remaining in the cartridge
will be displayed on the Home Screen
The amount of insulin remaining displayed
on the Home Screen will decrease 5 units
at a time (for example, you will see 140,
135, 130, 125). When less than 40 units
remain, it will begin decreasing 1 unit at a
time (for example, you will see 40, 39, 38,
37) until there is 1 unit remaining.
Summary of Contents for t:slim X2
Page 2: ......
Page 14: ...Section 1 Before You Begin...
Page 15: ...Chapter 1 Introduction...
Page 19: ...Chapter 2 Important Safety Information...
Page 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profiles...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim X2 Pump Settings...
Page 109: ...Chapter 12 t slim X2 Pump Info and History...
Page 111: ...Chapter 13 t slim X2 Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim X2 Pump Alerts...
Page 143: ...Chapter 16 t slim X2 Pump Alarms...
Page 157: ...Chapter 17 t slim X2 Pump Malfunction...
Page 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Page 161: ...Chapter 18 Lifestyle Issues and Travel...
Page 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Page 168: ...Section 6 Technical Specifications and Warranty...